*** MIDAS logfile (05SEPpl1.2) *** Sa, 06 May 2006 16:47:50 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 06 May 2006 16:48:47 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060506 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.84 stddev = 5.26 FIERA temperature = 289.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060506 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1848.0 1847.3 1528.3 -0.07 -0.08 1.21 2 2.13 3584.8 3584.2 2936.4 0.18 0.17 1.22 3 4.25 7027.5 7026.8 7263.8 -0.10 -0.08 0.97 4 6.38 10488.6 10487.2 8499.2 -0.03 -0.04 1.23 5 9.57 15673.2 15674.0 12936.5 -0.01 -0.01 1.21 6 12.76 20879.9 20881.6 16812.9 0.08 0.09 1.24 7 15.95 26069.7 26068.9 20743.7 0.09 0.09 1.26 8 21.27 34712.2 34711.3 27473.1 0.06 0.06 1.26 9 26.59 43326.1 43326.8 34081.8 -0.01 -0.01 1.27 10 31.91 51877.0 51875.7 40294.6 -0.19 -0.19 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2403 to 67480 e- Average count rate measured: 1625 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.02 e-/ADU RON = 6.84 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.077 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay