*** MIDAS logfile (05SEPpl1.2) *** Fr, 21 Apr 2006 17:36:45 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 21 Apr 2006 17:37:39 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060421 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.80 stddev = 5.31 FIERA temperature = 290.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060421 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1846.9 1847.1 1527.1 -0.04 -0.05 1.21 2 2.13 3589.3 3588.7 3889.9 0.20 0.18 0.92 3 4.25 7043.1 7041.8 5801.1 -0.08 -0.09 1.21 4 6.38 10512.5 10512.6 8582.9 -0.03 -0.03 1.22 5 9.57 15712.0 15712.3 13156.7 -0.02 -0.01 1.19 6 12.76 20933.5 20933.2 16896.3 0.09 0.09 1.24 7 15.96 26135.3 26135.9 20874.5 0.08 0.08 1.25 8 21.28 34798.3 34799.3 27411.1 0.05 0.05 1.27 9 26.59 43432.9 43433.2 34002.3 -0.03 -0.03 1.28 10 31.91 52009.3 52010.0 40081.6 -0.20 -0.20 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2417 to 68048 e- Average count rate measured: 1629 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.01 e-/ADU RON = 6.95 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.073 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.20 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay