*** MIDAS logfile (05SEPpl1.2) *** Fr, 17 Mar 2006 21:03:59 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 17 Mar 2006 21:05:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060317 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 322.42 stddev = 5.65 FIERA temperature = 289.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060317 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1801.6 1800.5 1491.0 -0.05 -0.09 1.21 2 2.13 3531.3 3530.7 2895.4 0.21 0.21 1.22 3 4.25 6955.2 6954.8 5692.8 -0.09 -0.09 1.22 4 6.38 10398.2 10398.0 8498.7 -0.03 -0.03 1.22 5 9.57 15555.9 15555.8 12675.8 -0.02 -0.01 1.23 6 12.76 20734.7 20733.3 17909.1 0.08 0.08 1.16 7 15.95 25893.6 25893.1 20916.2 0.07 0.08 1.24 8 21.27 34491.0 34491.0 27333.2 0.05 0.06 1.26 9 26.59 43057.1 43056.4 33762.4 -0.03 -0.02 1.28 10 31.91 51567.2 51566.1 39931.5 -0.19 -0.19 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2323 to 66522 e- Average count rate measured: 1616 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.29 +/- 0.02 e-/ADU RON = 7.29 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.055 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.21/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay