*** MIDAS logfile (05SEPpl1.2) *** Mo, 20 Feb 2006 19:21:57 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 20 Feb 2006 19:22:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060220 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 322.50 stddev = 5.51 FIERA temperature = 290.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060220 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1877.1 1876.6 1557.8 -0.06 -0.05 1.20 2 2.13 3644.9 3644.3 2994.4 0.19 0.17 1.22 3 4.25 7148.8 7149.8 6710.7 -0.09 -0.07 1.07 4 6.38 10671.3 10670.1 8736.0 -0.02 -0.03 1.22 5 9.57 15947.4 15947.0 12814.5 -0.01 -0.01 1.24 6 12.76 21244.6 21243.5 17199.5 0.09 0.09 1.24 7 15.96 26522.5 26521.9 21367.3 0.08 0.08 1.24 8 21.28 35312.4 35310.9 28120.9 0.05 0.05 1.26 9 26.59 44072.3 44073.7 34432.2 -0.03 -0.03 1.28 10 31.91 52770.6 52771.0 40613.7 -0.21 -0.21 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2444 to 68725 e- Average count rate measured: 1653 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.01 e-/ADU RON = 7.17 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.075 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay