*** MIDAS logfile (05SEPpl1.2) *** Mo, 13 Feb 2006 20:13:50 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 13 Feb 2006 20:14:48 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060213 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 321.42 stddev = 5.65 FIERA temperature = 289.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060213 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 940.7 941.1 803.8 0.05 0.05 1.17 2 2.00 1811.2 1811.5 1555.7 -0.09 -0.08 1.16 3 4.00 3555.7 3555.8 2928.5 -0.08 -0.09 1.21 4 6.00 5301.4 5301.2 4328.9 -0.06 -0.06 1.22 5 9.00 7918.4 7919.1 6470.1 -0.05 -0.05 1.22 6 12.00 10539.9 10540.0 8582.8 -0.00 -0.00 1.23 7 15.00 13158.6 13158.7 10685.5 0.01 0.01 1.23 8 20.00 17528.2 17530.3 14245.1 0.04 0.05 1.23 9 25.00 21900.3 21900.7 17589.4 0.08 0.08 1.25 10 30.00 26273.2 26273.4 21266.1 0.10 0.10 1.24 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1169 to 32641 e- Average count rate measured: 873 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.24 +/- 0.00 e-/ADU RON = 7.02 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.081 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay