*** MIDAS logfile (05SEPpl1.2) *** Sa, 11 Feb 2006 12:14:41 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 11 Feb 2006 12:15:38 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060211 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 320.98 stddev = 5.29 FIERA temperature = 289.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060211 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 956.3 955.4 884.7 0.07 0.06 1.08 2 2.00 1834.9 1834.3 1525.5 -0.08 -0.09 1.20 3 4.00 3593.3 3592.8 3000.6 -0.09 -0.10 1.20 4 6.00 5354.1 5354.3 4584.2 -0.06 -0.04 1.17 5 9.00 7992.4 7994.0 6512.6 -0.06 -0.04 1.23 6 12.00 10637.2 10636.2 8796.0 -0.00 -0.01 1.21 7 15.00 13280.1 13278.6 10835.9 0.02 0.01 1.23 8 20.00 17685.6 17685.3 14420.4 0.04 0.05 1.23 9 25.00 22093.4 22094.2 17908.3 0.07 0.08 1.23 10 30.00 26501.1 26503.0 21249.5 0.08 0.09 1.25 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1193 to 33079 e- Average count rate measured: 880 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.25 +/- 0.01 e-/ADU RON = 6.60 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.089 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay