*** MIDAS logfile (05SEPpl1.2) *** Fr, 27 Jan 2006 15:32:31 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 27 Jan 2006 15:33:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060127 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 326.97 stddev = 4.43 FIERA temperature = 291.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060127 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 946.4 946.2 794.7 0.04 0.07 1.19 2 2.00 1827.8 1827.4 1558.2 0.02 -0.00 1.17 3 4.00 3586.0 3586.2 2945.2 -0.08 -0.08 1.22 4 6.00 5347.1 5347.5 4396.3 -0.05 -0.05 1.22 5 9.00 7989.1 7989.6 6535.4 -0.04 -0.03 1.22 6 12.00 10631.6 10632.5 8622.2 -0.02 -0.02 1.23 7 15.00 13275.7 13275.8 10782.0 0.00 -0.00 1.23 8 20.00 17683.0 17683.1 14349.0 0.03 0.02 1.23 9 25.00 22090.9 22091.2 17830.6 0.05 0.04 1.24 10 30.00 26501.4 26502.3 21677.8 0.06 0.06 1.22 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1166 to 32663 e- Average count rate measured: 881 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.23 +/- 0.00 e-/ADU RON = 5.46 +/- 0.02 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.078 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.08 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay