*** MIDAS logfile (05SEPpl1.2) *** Sa, 21 Jan 2006 16:16:01 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 21 Jan 2006 16:17:06 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060121 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 323.61 stddev = 5.92 FIERA temperature = 291.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060121 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 932.3 931.8 796.5 0.01 0.06 1.17 2 2.00 1823.0 1823.3 1542.1 -0.07 -0.03 1.18 3 4.00 3606.6 3606.0 2987.6 -0.03 -0.03 1.21 4 6.00 5389.2 5389.1 4416.3 -0.03 -0.04 1.22 5 9.00 8063.1 8063.0 6778.8 -0.04 -0.04 1.19 6 12.00 10739.5 10739.7 8875.1 -0.01 -0.02 1.21 7 15.00 13416.8 13417.7 10970.3 0.01 0.01 1.22 8 20.00 17878.6 17879.9 14444.8 0.03 0.02 1.24 9 25.00 22342.8 22342.0 18040.8 0.05 0.03 1.24 10 30.00 26806.0 26804.9 21558.7 0.06 0.04 1.24 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1162 to 33423 e- Average count rate measured: 892 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.25 +/- 0.01 e-/ADU RON = 7.38 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.047 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.06/-0.05 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay