*** MIDAS logfile (05SEPpl1.2) *** Sa, 14 Jan 2006 19:46:09 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 14 Jan 2006 19:47:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 060114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 317.53 stddev = 4.10 FIERA temperature = 291.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_060114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 940.5 939.8 902.8 0.06 0.08 1.04 2 2.00 1828.3 1828.8 1518.7 -0.14 -0.10 1.20 3 4.00 3610.1 3609.7 2991.2 -0.06 -0.06 1.21 4 6.00 5390.3 5390.9 4420.9 -0.05 -0.04 1.22 5 9.00 8061.1 8061.1 6579.7 -0.03 -0.04 1.23 6 12.00 10734.4 10735.1 9009.1 -0.00 -0.01 1.19 7 15.00 13409.2 13410.0 10988.5 0.02 0.01 1.22 8 20.00 17865.6 17866.1 14546.5 0.04 0.03 1.23 9 25.00 22326.1 22326.0 18031.6 0.08 0.06 1.24 10 30.00 26783.8 26784.3 21495.4 0.08 0.07 1.25 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1172 to 33403 e- Average count rate measured: 890 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.25 +/- 0.01 e-/ADU RON = 5.11 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.058 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay