*** MIDAS logfile (03SEPpl1.2) *** Fr, 04 Nov 2005 18:25:59 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 04 Nov 2005 18:27:09 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 051104 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 346.07 stddev = 3.89 FIERA temperature = 291.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_051104 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 962.3 962.6 808.4 0.05 0.04 1.19 2 2.00 1854.1 1853.8 1537.8 -0.06 -0.08 1.21 3 4.00 3639.3 3639.2 2991.1 -0.06 -0.06 1.22 4 6.00 5424.2 5424.0 4427.9 -0.06 -0.06 1.22 5 9.00 8103.5 8103.8 6583.7 -0.04 -0.04 1.23 6 12.00 10783.8 10783.0 8751.6 -0.02 -0.02 1.23 7 15.00 13464.7 13466.3 10938.4 0.00 0.02 1.23 8 20.00 17937.5 17936.9 14652.4 0.05 0.05 1.22 9 25.00 22407.1 22409.7 18161.4 0.06 0.07 1.23 10 30.00 26879.2 26878.6 21683.5 0.08 0.08 1.24 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1191 to 33275 e- Average count rate measured: 893 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.24 +/- 0.00 e-/ADU RON = 4.81 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.081 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.07 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay