*** MIDAS logfile (03SEPpl1.2) *** Fr, 20 May 2005 19:04:09 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 20 May 2005 19:05:03 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 050520 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 345.97 stddev = 7.15 FIERA temperature = 290.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_050520 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 986.7 986.7 855.8 0.03 0.01 1.15 2 2.00 1898.7 1898.3 1643.1 -0.04 -0.05 1.16 3 4.00 3722.7 3722.3 3100.0 -0.06 -0.06 1.20 4 6.00 5547.0 5546.5 4694.0 -0.06 -0.06 1.18 5 9.00 8285.7 8285.8 6777.1 -0.04 -0.03 1.22 6 12.00 11026.0 11025.2 9004.9 -0.01 -0.01 1.22 7 15.00 13765.7 13765.5 11224.7 0.01 0.01 1.23 8 20.00 18335.9 18337.3 16046.9 0.04 0.05 1.14 9 25.00 22905.5 22906.2 18588.8 0.06 0.07 1.23 10 30.00 27474.0 27475.0 23009.8 0.07 0.08 1.19 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1184 to 32972 e- Average count rate measured: 913 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.20 +/- 0.02 e-/ADU RON = 8.58 +/- 0.13 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.083 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay