*** MIDAS logfile (03SEPpl1.2) *** Sa, 01 Jan 2005 17:29:33 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Sa, 01 Jan 2005 17:30:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC2 TAKEN ON 050101 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 326.34 stddev = 6.04 FIERA temperature = 291.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC2_050101 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1001.8 1002.1 857.5 0.05 0.07 1.17 2 2.00 1931.2 1930.7 1602.2 -0.06 -0.07 1.21 3 4.00 3788.5 3789.0 3102.6 -0.10 -0.09 1.22 4 6.00 5649.5 5649.0 4667.6 -0.07 -0.07 1.21 5 9.00 8439.9 8439.3 6906.7 -0.04 -0.05 1.22 6 12.00 11232.4 11232.5 9157.7 -0.01 -0.02 1.23 7 15.00 14026.4 14025.0 11665.8 0.02 0.01 1.20 8 20.00 18684.5 18683.7 15250.7 0.05 0.05 1.23 9 25.00 23342.1 23342.8 18749.9 0.08 0.08 1.24 10 30.00 28001.9 28002.9 22550.8 0.10 0.10 1.24 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1246 to 34812 e- Average count rate measured: 930 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.24 +/- 0.01 e-/ADU RON = 7.51 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.079 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay