*** MIDAS logfile (05SEPpl1.2) *** Sa, 09 Feb 2008 15:48:44 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 09 Feb 2008 15:49:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 080209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 230.79 stddev = 4.63 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_080209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1490.7 1491.0 1266.0 0.04 0.04 1.18 2 8.51 2965.7 2965.6 2394.5 -0.02 -0.02 1.24 3 17.02 5912.7 5912.6 4690.8 -0.09 -0.09 1.26 4 25.53 8862.0 8862.0 6988.4 -0.09 -0.09 1.27 5 38.30 13293.5 13293.2 10374.7 -0.03 -0.03 1.28 6 51.06 17724.9 17724.7 13666.0 0.00 0.00 1.30 7 63.83 22162.9 22162.9 16832.6 0.05 0.05 1.32 8 85.10 29554.9 29555.2 22061.8 0.08 0.08 1.34 9 106.38 36932.9 36933.2 26984.5 0.06 0.06 1.37 10 127.66 44289.3 44288.4 31489.3 -0.00 -0.00 1.41 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2091 to 62124 e- Average count rate measured: 347 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.40 +/- 0.02 e-/ADU RON = 6.49 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.045 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay