*** MIDAS logfile (05SEPpl1.2) *** We, 30 Jan 2008 16:35:41 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 30 Jan 2008 16:35:58 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 080130 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 220.84 stddev = 4.82 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_080130 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1517.7 1517.0 1244.6 0.04 0.04 1.22 2 8.51 3009.7 3009.5 2419.9 -0.01 -0.02 1.24 3 17.02 5990.2 5990.2 4751.4 -0.10 -0.10 1.26 4 25.53 8974.5 8975.0 7075.7 -0.09 -0.08 1.27 5 38.29 13457.9 13457.9 10514.0 -0.03 -0.03 1.28 6 51.06 17940.7 17941.2 13818.8 0.00 0.00 1.30 7 63.83 22429.0 22430.3 17082.0 0.04 0.05 1.31 8 85.10 29910.2 29909.9 22220.3 0.09 0.08 1.35 9 106.38 37371.3 37369.3 27163.7 0.06 0.05 1.38 10 127.66 44814.8 44813.3 32654.3 -0.00 -0.00 1.37 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2102 to 62076 e- Average count rate measured: 351 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.39 +/- 0.01 e-/ADU RON = 6.68 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.071 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.09/-0.10 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay