*** MIDAS logfile (05SEPpl1.2) *** Th, 03 Jan 2008 15:12:53 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 03 Jan 2008 15:13:11 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 080103 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 232.09 stddev = 4.97 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_080103 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1503.4 1503.2 1314.5 0.05 0.04 1.14 2 8.51 2990.9 2990.6 2416.2 -0.02 -0.03 1.24 3 17.02 5962.8 5962.9 4739.3 -0.10 -0.09 1.26 4 25.53 8937.8 8937.3 7070.0 -0.09 -0.09 1.26 5 38.29 13407.3 13407.8 10456.0 -0.03 -0.02 1.28 6 51.06 17877.1 17876.8 13769.6 0.00 0.00 1.30 7 63.83 22352.6 22353.7 17015.7 0.04 0.05 1.31 8 85.10 29810.1 29810.0 22255.6 0.08 0.09 1.34 9 106.38 37249.4 37249.6 27029.1 0.06 0.06 1.38 10 127.66 44668.5 44668.3 31903.5 -0.01 -0.00 1.40 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2109 to 62677 e- Average count rate measured: 350 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.40 +/- 0.02 e-/ADU RON = 6.97 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.045 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay