*** MIDAS logfile (05SEPpl1.2) *** Th, 22 Nov 2007 13:46:22 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 22 Nov 2007 13:46:40 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 071122 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 224.50 stddev = 4.48 FIERA temperature = 283.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_071122 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1507.5 1507.6 1225.8 0.04 0.04 1.23 2 8.51 3004.4 3004.1 2417.7 -0.02 -0.03 1.24 3 17.02 5994.1 5993.7 4782.1 -0.09 -0.09 1.25 4 25.53 8987.3 8987.0 7117.9 -0.08 -0.08 1.26 5 38.30 13483.2 13483.6 10543.9 -0.03 -0.02 1.28 6 51.06 17979.9 17980.4 13891.9 0.00 0.01 1.29 7 63.83 22482.4 22482.6 17111.9 0.05 0.05 1.31 8 85.10 29983.5 29984.3 22315.0 0.08 0.09 1.34 9 106.38 37467.6 37467.1 27202.6 0.06 0.06 1.38 10 127.66 44928.8 44929.1 32186.9 -0.01 -0.01 1.40 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2110 to 62897 e- Average count rate measured: 352 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.40 +/- 0.02 e-/ADU RON = 6.28 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.031 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay