*** MIDAS logfile (05SEPpl1.2) *** We, 14 Nov 2007 16:42:23 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 14 Nov 2007 16:42:40 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 071114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 229.12 stddev = 4.51 FIERA temperature = 282.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_071114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1552.7 1552.8 1261.8 0.04 0.04 1.23 2 8.51 3077.4 3076.9 2477.8 -0.02 -0.03 1.24 3 17.02 6123.7 6123.5 4887.1 -0.08 -0.09 1.25 4 25.53 9172.5 9172.7 7266.4 -0.08 -0.08 1.26 5 38.30 13752.9 13752.7 10719.1 -0.02 -0.02 1.28 6 51.06 18333.6 18333.8 14306.4 0.01 0.01 1.28 7 63.83 22919.8 22919.7 17399.7 0.05 0.05 1.32 8 85.10 30559.2 30558.9 22762.4 0.08 0.08 1.34 9 106.38 38180.7 38181.2 27623.7 0.05 0.05 1.38 10 127.66 45778.1 45777.9 32354.0 -0.02 -0.02 1.41 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2194 to 64680 e- Average count rate measured: 358 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.41 +/- 0.02 e-/ADU RON = 6.38 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.079 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.08/-0.09 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay