*** MIDAS logfile (05SEPpl1.2) *** We, 19 Sep 2007 17:54:55 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 19 Sep 2007 17:55:13 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 070919 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 266.50 stddev = 4.55 FIERA temperature = 282.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_070919 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1587.7 1587.5 1313.8 -0.00 -0.00 1.21 2 8.51 3145.4 3145.3 2562.2 0.01 0.02 1.23 3 17.02 6256.7 6256.8 5054.8 -0.03 -0.03 1.24 4 25.53 9370.0 9370.3 7551.7 -0.02 -0.02 1.24 5 38.29 14045.1 14045.1 11057.9 0.02 0.02 1.27 6 51.06 18717.8 18718.2 14538.7 0.03 0.03 1.29 7 63.83 23395.5 23395.6 17968.2 0.06 0.05 1.30 8 85.10 31184.1 31183.6 23632.1 0.06 0.05 1.32 9 106.38 38944.3 38944.7 28388.7 -0.01 -0.02 1.37 10 127.66 46678.0 46679.7 33218.0 -0.12 -0.12 1.41 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2225 to 65420 e- Average count rate measured: 366 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.40 +/- 0.02 e-/ADU RON = 6.37 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.089 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.06/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay