*** MIDAS logfile (05SEPpl1.2) *** Su, 01 Jul 2007 19:04:25 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 01 Jul 2007 19:04:43 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 070701 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.10 stddev = 5.53 FIERA temperature = 284.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_070701 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1621.8 1622.1 1335.7 -0.00 -0.00 1.21 2 8.51 3214.4 3214.3 2646.7 0.02 0.02 1.21 3 17.02 6394.6 6394.8 5182.7 -0.02 -0.02 1.23 4 25.53 9577.5 9577.2 7649.5 -0.02 -0.02 1.25 5 38.30 14357.1 14357.4 11300.9 0.03 0.03 1.27 6 51.06 19134.1 19134.9 14880.3 0.04 0.04 1.29 7 63.83 23915.5 23916.4 18331.1 0.06 0.06 1.30 8 85.10 31874.1 31874.1 23876.6 0.05 0.05 1.33 9 106.38 39806.9 39807.1 28950.3 -0.02 -0.02 1.38 10 127.66 47707.7 47707.2 33931.0 -0.13 -0.14 1.41 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2281 to 67080 e- Average count rate measured: 374 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.41 +/- 0.02 e-/ADU RON = 7.77 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.085 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.06/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay