*** MIDAS logfile (05SEPpl1.2) *** Su, 20 Aug 2006 18:06:39 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 20 Aug 2006 18:07:57 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 060820 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 227.96 stddev = 4.62 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_060820 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1646.2 1646.1 1321.6 0.07 0.06 1.25 2 8.51 3278.9 3279.2 2639.4 -0.04 -0.04 1.24 3 17.02 6542.4 6541.8 5165.5 -0.12 -0.12 1.27 4 25.53 9809.6 9810.2 7725.1 -0.10 -0.10 1.27 5 38.29 14719.7 14720.0 11419.9 -0.03 -0.03 1.29 6 51.06 19631.2 19631.5 14999.7 0.01 0.01 1.31 7 63.83 24548.9 24548.6 18612.7 0.06 0.06 1.32 8 85.10 32742.6 32743.4 24057.5 0.10 0.11 1.36 9 106.38 40908.3 40908.6 29316.0 0.06 0.06 1.40 10 127.66 49051.4 49050.8 34222.2 -0.01 -0.01 1.43 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2355 to 70160 e- Average count rate measured: 384 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.43 +/- 0.02 e-/ADU RON = 6.60 +/- 0.10 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.030 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.11/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay