*** MIDAS logfile (05SEPpl1.2) *** Sa, 05 Aug 2006 16:03:45 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 05 Aug 2006 16:04:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 060805 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 227.03 stddev = 5.42 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_060805 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1677.8 1678.1 1383.1 0.07 0.07 1.21 2 8.51 3326.6 3326.4 2661.4 -0.04 -0.05 1.25 3 17.02 6621.8 6621.9 5234.6 -0.12 -0.12 1.27 4 25.53 9921.3 9921.7 7759.9 -0.10 -0.10 1.28 5 38.30 14879.9 14879.8 11507.3 -0.03 -0.03 1.29 6 51.06 19839.4 19839.1 15172.8 0.01 0.01 1.31 7 63.83 24806.4 24806.4 18635.4 0.07 0.07 1.33 8 85.10 33078.1 33078.7 24345.4 0.10 0.11 1.36 9 106.38 41323.4 41324.6 29587.4 0.06 0.06 1.40 10 127.66 49544.1 49543.6 34537.0 -0.02 -0.02 1.43 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2402 to 70924 e- Average count rate measured: 388 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.43 +/- 0.02 e-/ADU RON = 7.77 +/- 0.12 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.070 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay