*** MIDAS logfile (05SEPpl1.2) *** We, 05 Jul 2006 17:40:38 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 05 Jul 2006 17:42:06 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 060705 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 227.14 stddev = 5.01 FIERA temperature = 288.4 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_060705 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1665.2 1665.2 1344.7 0.06 0.06 1.24 2 8.51 3314.3 3314.2 2658.4 -0.05 -0.05 1.25 3 17.02 6610.8 6610.5 5217.6 -0.11 -0.12 1.27 4 25.53 9910.5 9910.7 7828.6 -0.10 -0.10 1.27 5 38.30 14869.9 14869.8 11675.1 -0.03 -0.03 1.27 6 51.06 19830.7 19830.5 15100.7 0.01 0.01 1.31 7 63.83 24798.4 24799.1 18690.2 0.07 0.07 1.33 8 85.10 33072.2 33072.8 24384.5 0.10 0.11 1.36 9 106.38 41322.0 41322.9 29606.8 0.07 0.07 1.40 10 127.66 49541.1 49539.8 34382.6 -0.02 -0.02 1.44 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2390 to 71100 e- Average count rate measured: 388 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.44 +/- 0.02 e-/ADU RON = 7.18 +/- 0.12 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.036 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.10/-0.12 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay