*** MIDAS logfile (05SEPpl1.2) *** Fr, 19 May 2006 20:29:18 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 19 May 2006 20:30:30 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 060519 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.69 stddev = 4.57 FIERA temperature = 289.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_060519 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1700.0 1700.1 1402.8 -0.00 0.00 1.21 2 8.51 3387.7 3387.3 2756.3 0.01 -0.00 1.23 3 17.02 6759.6 6759.3 5426.2 -0.03 -0.03 1.25 4 25.53 10133.7 10133.9 8084.4 -0.02 -0.02 1.25 5 38.30 15202.3 15202.5 11905.0 0.03 0.04 1.28 6 51.06 20268.1 20267.2 15697.2 0.05 0.05 1.29 7 63.83 25337.7 25337.3 19364.1 0.07 0.07 1.31 8 85.10 33775.2 33773.8 25098.1 0.06 0.06 1.35 9 106.38 42179.9 42180.0 30496.1 -0.02 -0.02 1.38 10 127.66 50546.9 50546.9 35375.0 -0.15 -0.15 1.43 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2420 to 71941 e- Average count rate measured: 396 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.42 +/- 0.02 e-/ADU RON = 6.50 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.036 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay