*** MIDAS logfile (05SEPpl1.2) *** Sa, 21 Jan 2006 15:35:19 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 21 Jan 2006 15:36:39 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 060121 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 305.33 stddev = 7.16 FIERA temperature = 291.5 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_060121 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 894.6 894.7 779.2 0.04 0.04 1.15 2 8.00 1779.8 1779.8 1516.3 -0.05 -0.05 1.17 3 16.00 3551.3 3551.4 2911.0 -0.07 -0.07 1.22 4 24.00 5323.7 5323.6 4458.4 -0.06 -0.06 1.19 5 36.00 7982.9 7983.0 6425.0 -0.04 -0.04 1.24 6 48.00 10644.4 10644.5 8497.8 -0.01 -0.01 1.25 7 60.00 13306.2 13306.7 10664.6 0.01 0.01 1.25 8 80.00 17744.6 17745.0 14018.3 0.04 0.04 1.27 9 100.00 22184.8 22185.2 17074.0 0.07 0.07 1.30 10 120.00 26622.4 26622.1 20319.5 0.08 0.07 1.31 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1175 to 34968 e- Average count rate measured: 222 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.01 e-/ADU RON = 9.41 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.037 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.07 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay