*** MIDAS logfile (05SEPpl1.2) *** Sa, 14 Jan 2006 17:47:00 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 14 Jan 2006 17:48:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rC1 TAKEN ON 060114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 258.46 stddev = 3.92 FIERA temperature = 291.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rC1_060114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.00 897.6 897.4 822.6 0.04 0.04 1.09 2 8.00 1782.8 1782.5 1455.6 -0.04 -0.04 1.22 3 16.00 3553.4 3553.3 2892.3 -0.06 -0.06 1.23 4 24.00 5324.5 5324.3 4517.0 -0.06 -0.06 1.18 5 36.00 7982.3 7982.1 6434.3 -0.04 -0.04 1.24 6 48.00 10642.3 10642.2 8455.5 -0.02 -0.01 1.26 7 60.00 13303.2 13303.2 10501.2 0.01 0.01 1.27 8 80.00 17738.5 17739.2 13829.5 0.03 0.04 1.28 9 100.00 22175.9 22176.4 17100.0 0.06 0.06 1.30 10 120.00 26612.4 26611.8 20189.1 0.07 0.07 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1186 to 35165 e- Average count rate measured: 222 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.01 e-/ADU RON = 5.18 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.052 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.07/-0.06 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay