*** MIDAS logfile (05SEPpl1.2) *** Su, 02 Mar 2008 16:46:43 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 02 Mar 2008 16:46:58 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 080302 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 305.59 stddev = 5.70 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_080302 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1809.2 1808.9 2336.8 0.10 0.05 0.77 2 2.12 3515.7 3517.6 2919.1 0.06 0.08 1.20 3 4.25 6904.9 6905.1 5395.4 -0.30 -0.29 1.28 4 6.38 10312.1 10313.0 9239.9 -0.22 -0.21 1.12 5 9.57 15430.1 15429.2 12082.7 -0.14 -0.14 1.28 6 12.76 20573.9 20571.8 15936.7 0.02 0.02 1.29 7 15.95 25707.3 25705.0 20015.6 0.08 0.08 1.28 8 21.27 34262.8 34266.3 27180.4 0.14 0.15 1.26 9 26.59 42810.6 42812.8 32406.8 0.16 0.17 1.32 10 31.91 51312.7 51312.2 38471.7 0.08 0.09 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2435 to 69066 e- Average count rate measured: 1603 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.02 e-/ADU RON = 7.67 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.068 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.29 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay