*** MIDAS logfile (05SEPpl1.2) *** Sa, 08 Dec 2007 16:19:15 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 08 Dec 2007 16:19:29 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 071208 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 279.13 stddev = 5.40 FIERA temperature = 286.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_071208 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1874.8 1875.5 1490.7 0.06 0.09 1.26 2 2.12 3607.4 3608.1 2836.1 0.09 0.07 1.27 3 4.25 7047.4 7045.8 5628.9 -0.27 -0.28 1.25 4 6.38 10506.1 10506.7 8752.3 -0.21 -0.21 1.20 5 9.57 15698.1 15700.3 12179.9 -0.14 -0.13 1.29 6 12.76 20916.5 20917.8 16203.6 0.02 0.02 1.29 7 15.95 26123.8 26125.2 20059.2 0.08 0.08 1.30 8 21.27 34809.8 34809.2 27218.5 0.15 0.14 1.28 9 26.59 43479.5 43480.6 33007.6 0.16 0.16 1.32 10 31.91 52099.3 52099.3 38727.7 0.07 0.06 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2508 to 69694 e- Average count rate measured: 1627 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.01 e-/ADU RON = 7.23 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.092 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.28 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay