*** MIDAS logfile (05SEPpl1.2) *** We, 14 Nov 2007 17:10:59 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 14 Nov 2007 17:11:14 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 071114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 277.81 stddev = 4.92 FIERA temperature = 282.9 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_071114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1897.9 1896.5 1514.3 0.07 0.01 1.25 2 2.12 3640.7 3641.5 2847.4 0.11 0.13 1.28 3 4.25 7096.2 7096.1 5535.2 -0.28 -0.28 1.28 4 6.38 10574.2 10573.6 8243.7 -0.20 -0.20 1.28 5 9.57 15791.0 15791.5 12289.1 -0.14 -0.13 1.28 6 12.76 21038.3 21037.5 16361.0 0.03 0.03 1.29 7 15.95 26268.6 26268.5 20273.7 0.07 0.08 1.30 8 21.27 34998.2 34997.3 26847.1 0.14 0.15 1.30 9 26.59 43709.0 43711.6 33245.6 0.14 0.16 1.31 10 31.91 52374.5 52375.3 39238.2 0.06 0.07 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2522 to 69635 e- Average count rate measured: 1635 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 6.54 +/- 0.04 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.101 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.28 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay