*** MIDAS logfile (05SEPpl1.2) *** Th, 18 Oct 2007 17:10:40 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 18 Oct 2007 17:10:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 071018 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 344.80 stddev = 5.06 FIERA temperature = 286.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_071018 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1918.9 1919.6 1549.8 -0.04 -0.03 1.24 2 2.12 3694.1 3693.7 3820.2 0.15 0.13 0.97 3 4.25 7212.1 7211.5 5764.1 -0.17 -0.17 1.25 4 6.38 10750.9 10751.3 8697.9 -0.10 -0.10 1.24 5 9.57 16055.7 16058.4 12784.4 -0.05 -0.04 1.26 6 12.76 21384.2 21384.8 16785.7 0.07 0.07 1.27 7 15.95 26698.2 26699.0 21720.7 0.09 0.09 1.23 8 21.27 35549.2 35550.3 28927.4 0.09 0.10 1.23 9 26.59 44377.7 44381.0 34441.2 0.05 0.05 1.29 10 31.91 53150.1 53149.8 39631.5 -0.09 -0.09 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2537 to 70264 e- Average count rate measured: 1662 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.03 e-/ADU RON = 6.69 +/- 0.13 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.096 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.14/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay