*** MIDAS logfile (05SEPpl1.2) *** We, 03 Oct 2007 16:28:41 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 03 Oct 2007 16:28:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 071003 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 345.53 stddev = 5.58 FIERA temperature = 285.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_071003 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1938.0 1938.8 2175.8 -0.07 -0.01 0.89 2 2.12 3720.6 3721.7 2973.8 0.16 0.21 1.25 3 4.25 7245.6 7246.6 5757.9 -0.17 -0.17 1.26 4 6.38 10793.9 10794.7 8572.4 -0.11 -0.11 1.26 5 9.57 16115.3 16116.0 12706.8 -0.05 -0.05 1.27 6 12.76 21456.4 21457.8 17148.5 0.07 0.05 1.25 7 15.95 26784.8 26784.7 20966.6 0.09 0.07 1.28 8 21.27 35661.4 35664.5 27730.8 0.10 0.09 1.29 9 26.59 44512.8 44512.6 33954.9 0.05 0.03 1.31 10 31.91 53310.6 53310.7 40288.4 -0.08 -0.10 1.32 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2576 to 70848 e- Average count rate measured: 1667 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 7.41 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.104 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay