*** MIDAS logfile (05SEPpl1.2) *** Th, 27 Sep 2007 18:24:13 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 27 Sep 2007 18:24:28 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 070927 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 351.91 stddev = 5.65 FIERA temperature = 282.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_070927 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1902.1 1903.4 1528.4 -0.04 -0.00 1.24 2 2.12 3689.4 3689.9 2942.9 0.15 0.16 1.25 3 4.25 7235.8 7236.3 5771.9 -0.16 -0.17 1.25 4 6.38 10802.1 10801.1 8876.8 -0.09 -0.11 1.22 5 9.57 16150.4 16149.8 12813.9 -0.04 -0.05 1.26 6 12.76 21518.8 21520.0 16996.1 0.07 0.06 1.27 7 15.95 26873.1 26872.4 21700.6 0.09 0.08 1.24 8 21.27 35793.4 35795.2 27625.2 0.10 0.09 1.30 9 26.59 44688.0 44688.3 33951.2 0.04 0.03 1.32 10 31.91 53521.6 53525.5 40140.1 -0.11 -0.11 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2533 to 71255 e- Average count rate measured: 1675 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.02 e-/ADU RON = 7.52 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.076 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay