*** MIDAS logfile (05SEPpl1.2) *** We, 19 Sep 2007 18:24:11 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 19 Sep 2007 18:24:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 070919 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 343.97 stddev = 5.03 FIERA temperature = 282.3 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_070919 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1941.6 1942.6 1550.5 -0.04 -0.03 1.25 2 2.12 3730.9 3731.6 2976.9 0.23 0.25 1.25 3 4.25 7266.3 7265.7 5795.0 -0.16 -0.17 1.25 4 6.38 10822.4 10821.2 8561.1 -0.10 -0.11 1.26 5 9.57 16156.8 16156.6 12708.2 -0.06 -0.06 1.27 6 12.76 21511.9 21512.4 16905.7 0.06 0.06 1.27 7 15.95 26851.0 26850.4 21026.9 0.07 0.07 1.28 8 21.27 35749.1 35750.4 27594.8 0.08 0.08 1.30 9 26.59 44622.5 44622.5 34093.1 0.03 0.02 1.31 10 31.91 53439.9 53441.1 41137.9 -0.11 -0.11 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2539 to 69861 e- Average count rate measured: 1671 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.01 e-/ADU RON = 6.58 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.104 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.24/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay