*** MIDAS logfile (05SEPpl1.2) *** We, 05 Sep 2007 17:01:50 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 05 Sep 2007 17:02:05 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 070905 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 349.64 stddev = 5.28 FIERA temperature = 283.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_070905 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1913.7 1912.5 1537.5 -0.01 -0.05 1.24 2 2.12 3716.3 3716.1 2974.8 0.15 0.16 1.25 3 4.25 7287.8 7286.7 5781.1 -0.17 -0.18 1.26 4 6.38 10880.4 10881.8 9352.1 -0.10 -0.09 1.16 5 9.57 16270.6 16267.0 12973.8 -0.04 -0.05 1.25 6 12.76 21676.2 21677.4 17066.9 0.07 0.07 1.27 7 15.95 27072.2 27071.5 21168.3 0.09 0.09 1.28 8 21.27 36058.1 36056.5 27756.0 0.09 0.10 1.30 9 26.59 45014.4 45017.3 34341.4 0.03 0.04 1.31 10 31.91 53917.3 53914.2 40231.8 -0.11 -0.11 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2561 to 72185 e- Average count rate measured: 1687 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.01 e-/ADU RON = 7.07 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.075 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay