*** MIDAS logfile (05SEPpl1.2) *** We, 11 Apr 2007 14:50:04 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 11 Apr 2007 14:50:19 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 070411 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.31 stddev = 7.07 FIERA temperature = 282.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_070411 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 2006.0 2007.3 1673.1 -0.07 -0.04 1.20 2 2.12 3851.8 3851.5 3276.7 0.17 0.18 1.18 3 4.25 7503.1 7504.2 5949.4 -0.16 -0.15 1.26 4 6.38 11178.3 11177.3 8939.5 -0.08 -0.09 1.25 5 9.57 16687.1 16685.5 13249.0 -0.04 -0.05 1.26 6 12.76 22217.2 22217.6 17328.6 0.07 0.07 1.28 7 15.95 27735.4 27734.4 21630.4 0.10 0.09 1.28 8 21.27 36924.1 36923.1 28652.9 0.10 0.09 1.29 9 26.59 46080.4 46080.8 35094.3 0.03 0.03 1.31 10 31.91 55174.9 55174.3 41297.8 -0.12 -0.13 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2676 to 73579 e- Average count rate measured: 1726 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 9.43 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.104 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.17/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay