*** MIDAS logfile (05SEPpl1.2) *** Mo, 26 Mar 2007 17:09:27 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 26 Mar 2007 17:09:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 070326 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 354.65 stddev = 5.49 FIERA temperature = 285.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_070326 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1945.0 1946.3 1585.1 -0.05 -0.07 1.23 2 2.12 3804.4 3805.3 3998.2 0.18 0.18 0.95 3 4.25 7485.8 7484.4 6002.0 -0.15 -0.17 1.25 4 6.38 11188.1 11187.3 8862.0 -0.09 -0.09 1.26 5 9.57 16741.4 16740.9 13341.5 -0.03 -0.03 1.25 6 12.76 22314.6 22314.8 17577.6 0.08 0.08 1.27 7 15.95 27873.0 27873.6 21747.0 0.10 0.10 1.28 8 21.27 37130.8 37130.3 28476.1 0.09 0.10 1.30 9 26.59 46354.5 46355.3 35216.9 0.02 0.03 1.32 10 31.91 55518.6 55517.6 41446.7 -0.14 -0.13 1.34 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2622 to 74828 e- Average count rate measured: 1739 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.01 e-/ADU RON = 7.39 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.059 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.16 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay