*** MIDAS logfile (05SEPpl1.2) *** We, 14 Mar 2007 18:28:20 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 14 Mar 2007 18:28:35 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 070314 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.93 stddev = 5.54 FIERA temperature = 282.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_070314 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 2012.9 2012.5 1624.9 -0.02 -0.05 1.24 2 2.12 3877.7 3877.9 3103.6 0.18 0.18 1.25 3 4.25 7573.9 7574.2 6014.7 -0.17 -0.16 1.26 4 6.38 11290.3 11292.5 9084.1 -0.10 -0.08 1.24 5 9.57 16867.1 16865.5 13570.8 -0.04 -0.05 1.24 6 12.76 22466.2 22465.2 17670.7 0.08 0.08 1.27 7 15.95 28047.7 28048.1 21796.7 0.09 0.10 1.29 8 21.27 37345.8 37344.4 28703.3 0.09 0.09 1.30 9 26.59 46609.9 46612.5 35910.9 0.02 0.03 1.30 10 31.91 55812.1 55812.0 41897.5 -0.14 -0.13 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2670 to 74035 e- Average count rate measured: 1746 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.33 +/- 0.01 e-/ADU RON = 7.35 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.093 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay