*** MIDAS logfile (05SEPpl1.2) *** Su, 29 Oct 2006 16:39:48 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 29 Oct 2006 16:40:46 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 061029 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.55 stddev = 5.14 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_061029 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 2054.8 2055.0 1639.2 -0.01 -0.02 1.25 2 2.12 3957.2 3956.5 3133.1 0.14 0.11 1.26 3 4.25 7726.7 7727.1 6152.8 -0.20 -0.19 1.26 4 6.38 11520.3 11520.8 9047.8 -0.11 -0.11 1.27 5 9.57 17208.3 17207.1 13571.9 -0.06 -0.06 1.27 6 12.76 22922.1 22923.1 18460.1 0.08 0.08 1.24 7 15.95 28618.4 28619.5 22325.9 0.10 0.11 1.28 8 21.27 38111.0 38109.8 29246.3 0.11 0.11 1.30 9 26.59 47567.8 47567.7 36126.6 0.05 0.05 1.32 10 31.91 56960.6 56965.1 42326.4 -0.10 -0.09 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2751 to 76261 e- Average count rate measured: 1782 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.34 +/- 0.02 e-/ADU RON = 6.88 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.094 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.12/-0.19 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay