*** MIDAS logfile (05SEPpl1.2) *** We, 06 Sep 2006 15:54:08 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 06 Sep 2006 15:54:55 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 060906 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 277.23 stddev = 5.20 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_060906 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1987.4 1987.1 1553.6 0.12 0.06 1.28 2 2.12 3880.1 3878.6 3046.9 0.04 0.02 1.27 3 4.25 7635.0 7635.5 5934.5 -0.35 -0.34 1.29 4 6.38 11419.0 11417.8 8782.7 -0.25 -0.24 1.30 5 9.57 17099.2 17098.5 13287.9 -0.14 -0.13 1.29 6 12.76 22808.0 22806.5 17981.8 0.04 0.04 1.27 7 15.95 28506.5 28506.7 21945.1 0.11 0.12 1.30 8 21.27 38005.4 38007.5 29026.4 0.18 0.19 1.31 9 26.59 47481.5 47482.7 35869.8 0.18 0.19 1.32 10 31.91 56900.6 56902.3 41777.2 0.08 0.09 1.36 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2680 to 76734 e- Average count rate measured: 1778 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.02 e-/ADU RON = 7.01 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.056 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.34 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay