*** MIDAS logfile (05SEPpl1.2) *** Su, 20 Aug 2006 18:42:59 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Su, 20 Aug 2006 18:43:56 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 060820 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 276.72 stddev = 5.43 FIERA temperature = 288.2 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_060820 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 1999.5 2000.1 1564.8 0.08 0.07 1.28 2 2.13 3903.0 3903.7 3012.4 0.05 0.07 1.30 3 4.25 7678.0 7678.2 5988.4 -0.33 -0.34 1.28 4 6.38 11479.5 11480.3 8847.7 -0.25 -0.25 1.30 5 9.57 17190.0 17188.2 13329.6 -0.14 -0.14 1.29 6 12.76 22927.9 22927.7 17664.8 0.03 0.04 1.30 7 15.96 28656.7 28654.3 21954.6 0.11 0.11 1.31 8 21.27 38206.9 38205.7 29914.0 0.19 0.18 1.28 9 26.59 47729.2 47728.8 36015.3 0.18 0.18 1.33 10 31.91 57192.8 57195.3 43026.7 0.07 0.07 1.33 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2648 to 75720 e- Average count rate measured: 1788 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.32 +/- 0.01 e-/ADU RON = 7.18 +/- 0.06 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.057 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.34 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay