*** MIDAS logfile (05SEPpl1.2) *** Tu, 01 Aug 2006 17:24:46 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Tu, 01 Aug 2006 17:25:41 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 060801 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.27 stddev = 5.25 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_060801 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 2024.5 2026.2 1658.7 0.07 0.10 1.22 2 2.12 3925.4 3925.5 3065.8 0.06 0.06 1.28 3 4.25 7692.6 7691.2 5984.8 -0.33 -0.34 1.29 4 6.38 11484.9 11485.1 8931.2 -0.25 -0.25 1.29 5 9.57 17183.3 17181.4 13312.0 -0.14 -0.15 1.29 6 12.76 22908.2 22907.7 17649.1 0.04 0.03 1.30 7 15.95 28622.6 28623.6 22208.0 0.11 0.11 1.29 8 21.27 38151.5 38152.0 29417.5 0.18 0.18 1.30 9 26.59 47654.4 47657.5 35690.2 0.18 0.18 1.34 10 31.91 57099.7 57100.4 42070.5 0.07 0.07 1.36 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2731 to 76992 e- Average count rate measured: 1784 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.01 e-/ADU RON = 7.08 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.075 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.33 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay