*** MIDAS logfile (05SEPpl1.2) *** We, 12 Jul 2006 23:58:02 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** We, 12 Jul 2006 23:59:11 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 060712 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 275.63 stddev = 5.58 FIERA temperature = 288.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_060712 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 2024.0 2025.3 1615.3 0.10 0.09 1.25 2 2.12 3942.8 3942.9 3116.1 0.07 0.07 1.27 3 4.25 7746.2 7745.8 5974.7 -0.33 -0.33 1.30 4 6.38 11577.3 11578.9 9002.8 -0.24 -0.23 1.29 5 9.57 17329.0 17329.5 13394.7 -0.14 -0.14 1.29 6 12.76 23111.2 23110.3 17812.8 0.04 0.03 1.30 7 15.95 28879.4 28879.3 22489.1 0.10 0.10 1.28 8 21.27 38500.1 38500.7 29203.3 0.18 0.18 1.32 9 26.59 48093.1 48093.0 35900.4 0.17 0.17 1.34 10 31.91 57624.4 57625.4 42539.9 0.06 0.06 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2736 to 77871 e- Average count rate measured: 1801 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.01 e-/ADU RON = 7.55 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.063 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.18/-0.33 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay