*** MIDAS logfile (05SEPpl1.2) *** Th, 06 Apr 2006 17:26:30 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 06 Apr 2006 17:27:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 060406 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 353.00 stddev = 5.52 FIERA temperature = 289.7 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_060406 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 2147.1 2146.8 1716.6 -0.00 -0.02 1.25 2 2.13 4150.1 4148.1 3276.1 0.17 0.14 1.27 3 4.25 8114.7 8115.6 6403.7 -0.17 -0.16 1.27 4 6.38 12104.5 12105.1 9787.4 -0.09 -0.08 1.24 5 9.57 18086.8 18086.2 14929.4 -0.04 -0.04 1.21 6 12.76 24095.6 24094.3 18688.6 0.09 0.09 1.29 7 15.95 30085.4 30085.9 23388.2 0.11 0.12 1.29 8 21.27 40058.2 40059.3 31267.6 0.10 0.11 1.28 9 26.59 49988.5 49991.1 40449.0 0.01 0.02 1.24 10 31.91 59846.6 59847.8 44351.6 -0.17 -0.17 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2811 to 78366 e- Average count rate measured: 1873 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.03 e-/ADU RON = 7.23 +/- 0.15 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.086 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay