*** MIDAS logfile (05SEPpl1.2) *** Mo, 20 Feb 2006 19:18:55 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Mo, 20 Feb 2006 19:19:53 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 060220 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 350.06 stddev = 5.76 FIERA temperature = 290.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_060220 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.06 2140.1 2139.8 1767.2 -0.00 0.01 1.21 2 2.13 4137.2 4136.3 3312.5 0.14 0.12 1.25 3 4.25 8097.9 8097.8 6504.4 -0.17 -0.16 1.24 4 6.38 12081.7 12080.0 9595.8 -0.09 -0.10 1.26 5 9.57 18055.7 18054.6 14221.6 -0.04 -0.04 1.27 6 12.76 24056.0 24055.0 18810.5 0.09 0.10 1.28 7 15.96 30037.3 30035.1 23171.5 0.12 0.11 1.30 8 21.28 39995.3 39993.4 30634.8 0.10 0.11 1.31 9 26.59 49913.6 49912.2 37852.3 0.02 0.02 1.32 10 31.91 59754.7 59753.3 44164.3 -0.17 -0.16 1.35 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2882 to 80475 e- Average count rate measured: 1871 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.35 +/- 0.01 e-/ADU RON = 7.76 +/- 0.07 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.083 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.13/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay