*** MIDAS logfile (05SEPpl1.2) *** Fr, 27 Jan 2006 15:31:32 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Fr, 27 Jan 2006 15:32:26 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 060127 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 347.63 stddev = 4.68 FIERA temperature = 291.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_060127 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1089.6 1089.3 1009.7 0.08 0.11 1.08 2 2.00 2094.1 2094.0 1667.6 -0.03 -0.03 1.26 3 4.00 4099.2 4098.8 3271.1 -0.14 -0.15 1.25 4 6.00 6107.8 6108.7 5847.2 -0.12 -0.11 1.04 5 9.00 9122.6 9123.0 7938.1 -0.09 -0.08 1.15 6 12.00 12139.8 12141.0 9843.4 -0.04 -0.04 1.23 7 15.00 15160.2 15159.7 12001.4 -0.00 -0.01 1.26 8 20.00 20199.2 20197.8 15854.7 0.07 0.06 1.27 9 25.00 25236.2 25237.6 19973.0 0.11 0.11 1.26 10 30.00 30282.0 30282.4 23231.9 0.15 0.15 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1427 to 39652 e- Average count rate measured: 1005 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.31 +/- 0.02 e-/ADU RON = 6.12 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.087 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.15/-0.15 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay