*** MIDAS logfile (05SEPpl1.2) *** Sa, 14 Jan 2006 19:45:08 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 14 Jan 2006 19:46:04 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 060114 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 341.48 stddev = 4.07 FIERA temperature = 291.8 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_060114 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1104.7 1103.9 963.4 0.13 0.08 1.15 2 2.00 2134.1 2135.3 2021.2 -0.15 -0.11 1.06 3 4.00 4198.8 4198.9 3377.2 -0.13 -0.13 1.24 4 6.00 6263.9 6264.0 4944.3 -0.12 -0.12 1.27 5 9.00 9363.0 9363.3 7425.2 -0.09 -0.08 1.26 6 12.00 12467.3 12467.9 9791.6 -0.03 -0.03 1.27 7 15.00 15573.6 15574.3 12331.1 0.02 0.02 1.26 8 20.00 20754.7 20754.2 16231.8 0.08 0.08 1.28 9 25.00 25938.0 25938.8 20339.8 0.13 0.13 1.28 10 30.00 31120.7 31122.6 23856.6 0.16 0.16 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1439 to 40559 e- Average count rate measured: 1033 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.01 e-/ADU RON = 5.30 +/- 0.03 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.071 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay