*** MIDAS logfile (05SEPpl1.2) *** Th, 29 Dec 2005 19:22:18 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Th, 29 Dec 2005 19:23:15 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 051229 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 368.10 stddev = 5.70 FIERA temperature = 291.6 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_051229 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1092.5 1092.2 913.3 0.04 0.10 1.20 2 2.00 2097.4 2096.1 1718.2 -0.04 -0.05 1.22 3 4.00 4103.6 4105.0 3288.2 -0.12 -0.11 1.25 4 6.00 6111.6 6112.6 4812.8 -0.14 -0.13 1.27 5 9.00 9128.6 9130.2 7281.2 -0.08 -0.09 1.25 6 12.00 12148.4 12147.8 9582.0 -0.04 -0.05 1.27 7 15.00 15170.0 15169.6 12727.8 0.01 -0.01 1.19 8 20.00 20213.9 20213.1 16087.7 0.09 0.07 1.26 9 25.00 25255.2 25256.8 19787.3 0.13 0.11 1.28 10 30.00 30299.9 30302.6 23644.2 0.16 0.15 1.28 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1395 to 38691 e- Average count rate measured: 1006 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.28 +/- 0.02 e-/ADU RON = 7.27 +/- 0.09 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.089 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.13 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay