*** MIDAS logfile (03SEPpl1.2) *** We, 14 Sep 2005 17:40:55 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** We, 14 Sep 2005 17:42:02 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 050914 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 357.61 stddev = 4.06 FIERA temperature = 290.0 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_050914 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1129.6 1130.4 912.0 0.10 0.08 1.24 2 2.00 2179.6 2178.7 1730.8 -0.04 -0.10 1.26 3 4.00 4277.6 4276.1 3398.4 -0.13 -0.15 1.26 4 6.00 6377.3 6375.1 5030.7 -0.13 -0.13 1.27 5 9.00 9529.3 9530.1 7524.9 -0.09 -0.08 1.27 6 12.00 12686.0 12684.9 10000.0 -0.04 -0.03 1.27 7 15.00 15844.8 15844.8 12710.7 0.01 0.02 1.25 8 20.00 21113.7 21114.3 18058.1 0.07 0.09 1.17 9 25.00 26384.3 26385.9 20529.1 0.11 0.14 1.29 10 30.00 31656.7 31655.6 24510.8 0.15 0.17 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1435 to 40189 e- Average count rate measured: 1051 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.03 e-/ADU RON = 5.15 +/- 0.11 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.076 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.14 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay