*** MIDAS logfile (03SEPpl1.2) *** Fr, 04 Mar 2005 16:22:27 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Fr, 04 Mar 2005 16:23:54 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 050304 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 348.56 stddev = 6.82 FIERA temperature = 291.3 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_050304 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1136.2 1135.0 937.3 0.12 0.13 1.21 2 2.00 2179.7 2179.4 1745.5 -0.07 -0.05 1.25 3 4.00 4267.0 4266.4 3375.2 -0.17 -0.17 1.26 4 6.00 6357.8 6356.7 4985.4 -0.14 -0.15 1.28 5 9.00 9495.6 9495.1 7510.2 -0.11 -0.11 1.26 6 12.00 12637.3 12637.5 9929.0 -0.06 -0.06 1.27 7 15.00 15783.0 15782.7 12286.4 0.00 -0.01 1.28 8 20.00 21032.0 21033.6 16481.4 0.08 0.08 1.28 9 25.00 26283.9 26286.0 21214.8 0.15 0.15 1.24 10 30.00 31536.3 31536.2 24492.2 0.19 0.18 1.29 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1445 to 40123 e- Average count rate measured: 1046 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.27 +/- 0.01 e-/ADU RON = 8.68 +/- 0.08 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.086 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay