*** MIDAS logfile (03SEPpl1.2) *** Mo, 28 Feb 2005 17:45:29 Page 1 *** MIDAS logfile (vers. 03SEPpl1.2) *** Mo, 28 Feb 2005 17:46:42 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB2 TAKEN ON 050228 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 345.24 stddev = 5.87 FIERA temperature = 291.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB2_050228 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 1.00 1135.2 1135.0 969.0 0.14 0.14 1.17 2 2.00 2177.8 2177.8 1731.0 -0.07 -0.09 1.26 3 4.00 4262.5 4262.4 3370.5 -0.17 -0.17 1.26 4 6.00 6351.0 6349.9 5025.6 -0.15 -0.16 1.26 5 9.00 9483.9 9484.5 7465.0 -0.12 -0.11 1.27 6 12.00 12623.3 12622.7 9788.9 -0.06 -0.06 1.29 7 15.00 15764.6 15765.5 12282.9 -0.00 0.00 1.28 8 20.00 21008.9 21008.5 16253.2 0.09 0.09 1.29 9 25.00 26254.4 26255.5 20262.4 0.15 0.16 1.30 10 30.00 31501.6 31500.7 24312.3 0.19 0.19 1.30 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 1477 to 40985 e- Average count rate measured: 1045 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.30 +/- 0.00 e-/ADU RON = 7.64 +/- 0.01 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.087 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.19/-0.17 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay