*** MIDAS logfile (05SEPpl1.2) *** Sa, 09 Feb 2008 15:48:19 Page 1 *** MIDAS logfile (vers. 05SEPpl1.2) *** Sa, 09 Feb 2008 15:48:36 Page 2 ========================================================== SUMMARY STANDARD TEST CCD#rB1 TAKEN ON 080209 IN normcd MODE ========================================================== BIAS ==== 5 biases at start test: average = 235.11 stddev = 5.42 FIERA temperature = 288.1 K There were 0 hot pixels/columns detected TRANSFER FUNCTION AND LINEARITY DATA ==================================== Table : Std_normcd_rB1_080209 Sequence Exp_tim1 Med_cnt1 Med_cnt2 Variance Frc_cnt1 Frc_cnt2 el_adu1 -------- --------- --------- --------- --------- --------- --------- --------- 1 4.25 1723.7 1724.2 1452.0 0.13 0.14 1.19 2 8.51 3423.9 3423.8 2720.0 -0.12 -0.12 1.26 3 17.02 6824.9 6825.1 5315.8 -0.21 -0.21 1.28 4 25.53 10231.5 10231.6 7759.8 -0.19 -0.19 1.32 5 38.30 15355.0 15354.8 11497.7 -0.09 -0.09 1.34 6 51.06 20484.8 20483.7 15160.1 -0.01 -0.01 1.35 7 63.83 25623.3 25623.2 18758.1 0.08 0.07 1.37 8 85.10 34185.8 34186.6 24222.1 0.15 0.16 1.41 9 106.38 42731.3 42732.4 29541.7 0.16 0.16 1.45 10 127.66 51246.2 51248.3 34206.0 0.11 0.11 1.50 -------- --------- --------- --------- --------- --------- --------- --------- Dynamic range explored: from 2576 to 76578 e- Average count rate measured: 401 ADU/sec. CONV. FACTOR AND RON ==================== C.Fact = 1.49 +/- 0.03 e-/ADU RON = 8.09 +/- 0.14 e- RMS SHUTTER ERROR ============= Average shutter error derived from fractional count rate = 0.043 sec LINEARITY ERROR =============== Average linearity error after shutter correct. = + 0.16/-0.21 % These errors represent maximum deviation in % from the average count rate after correction for shutter delay